...
5) If you need to reinitialize the detector than follow the same steps as initialization in the Old Full Startup Procedures, except do not reload the JADE2 firmware.
...
If that still doesn't work you will need to fully reset IGRINS. Shut down the detector control computers as discussed in Old Full Shutdown Procedures, then shut down DSC and ICC. Unplug IGRINS power from the power socket in the dome floor. Wait a minute. Plug it back in. Go up to the computer rack and turn off the power to the arc and flat lamps, and check the Lakeshores and turn the heaters on as described in Lakeshore Temperature and Heater Settings. Push the power buttons on the detector control computers. Close the computer rack up and head back to the control room. Turn DSC back on. Turn ICC back on. Follow the Old Full Startup Procedures for reactivating the detector computers and starting up the software on the ICC. Hopefully this should work.
...
Close the DTP, SVC, and Housekeeping packages. Reload ibunny.py. Rerun "bash icsrun". If this doesn't solve the problem, close and reload the packages a second time. That should fix it.
11) SVC package keeps saying "Finder is not found" even though your target is in the A or B box when guiding:
...