9)Socket errors, ibunny.py not working, or SVC and DTP exposures not being taken? First try restarting the ICC and following the Starting up IGRINS computers & data taking software. If that doens't work try restarting the DSC and ICC and Starting up IGRINS computers & data taking software. If that doesn't work try shutting down the DSC & ICC, wait 5 minutes, then start them up again.
If If that still doesn't work you will need to fully reset IGRINS. Shut down the detector control computers as discussed in Full Shutdown Procedures, then shut down DSC and ICC. Unplug IGRINS power from the power socket in the dome floor. Wait a minute. Plug it back in. Go up to the computer rack and turn off the power to the arc and flat lamps, and check the Lakeshores and turn the heaters on as described in Lakeshore Temperature and Heater Settings. Push the power buttons on the detector control computers. Close the computer rack up and head back to the control room. Turn DSC back on. Turn ICC back on. Follow the Full Startup Procedures for reactivating the detector computers and starting up the software on the ICC. Hopefully this should work.
10) When trying to set up the guiding and clicking "Move A" or "Move B" does not work and you see "AG Failure" error in the DTP log and the SVC does not seem to display new images:
Close the DTP, SVC, and Housekeeping packages. Reload ibunny.py. Rerun "bash icsrun".