The labels on the Lakeshore are correctly labeled, and version 1.801 of the HKP is also correct.

 


The upper left Lakeshore 332 controls:

DET H: The H-band detector is on Channel A and should have a setpoint of 65K. Heater power should be set to mediumMED.

1ST 2ND Cold Head: The first cold head is the warmer of the two stages, is shown on Channel B, and has no heater or setpoint. Should say OFF.


The upper right Lakeshore 332 controls:

DET K: The K-band detector is on Channel A and should have a setpoint of 65K 66K. Heater power should be set to mediumMED.

Imm Gr: The immersion grating is on Channel B and should have a setpoint of 130K. Heater power should be set to mediumON.


The lower Lakeshore 350 controls:

Charcoal B: The charcoal box is shown on Channel D2, and has no heater or setpoint.

 

OPTICAL BENCH: The optical bench is shown on Channel A and should be set to 125K 120K. The heater power should be set to 5Rng5. 

DET SVC: The Slit View Camera detector is shown on Channel B and should have a setpoint of 65K 65.9K. Heater should be set to 5Rng5.2ND Cold Head: The second cold head is the coldest part of the IGRINS cryostat, is shown on Channel D3, and has no heater or setpoint.  


The rack temperatures should be between 273K and 320K. Typical temperatures are between 280K and 310K. The overheat shutoff for the electronics rack is 320K.