HEPA Filter Experiment Checklist
Operation Checklist
- Run dry air through the system to purge
- Run APS calibrationCharacterize salt through APS
- Remove filter
- Remove humidity/temp sensor
- Dilution 100:1 setting
- Set both pumps to 0.45 m/s flow rate
- Check diluter and total dP setting correct. adjust sheath flow rate if not
- Run dry air through system to purge
- Weigh clean filterBegin recording RH,TEMP, dP
- Install filter, mark orientation
- Place humidity sensor below filter
- Run dry air through system @ 0.28 m/s
- Load filter with salt
- Turn off blower and atomizer air
- Weigh filter
- Install filter at same orientation
- Place humidity sensor above filter
- Run air through humidifier @ 0.28 m/sFlow dry air through filter until dP constant (watch for sudden P drop, then steadies)
- Control humidifier to keep RH 78-79% for 5 min
- Run air through humidifier @ 0.28 m/sFlow dry air through filter until dP constant (watch for sudden P drop, then steadies)
- Control humidifier to keep RH 78-79% for 5 min
- Run air through humidifier @ 0.28 m/sFlow dry air through filter until dP constant (watch for sudden P drop, then steadies)
- Control humidifier to keep RH 78-79% for 5 min
- Turn off blower
- Weigh filter
Welcome to the University Wiki Service! Please use your IID (yourEID@eid.utexas.edu) when prompted for your email address during login or click here to enter your EID. If you are experiencing any issues loading content on pages, please try these steps to clear your browser cache.