Detectors not taking exposures

    Detectors not taking exposures

    Oct 17, 2014

     

    This problem usually manifests as 'Detector is not available' and 'Initialize was not completed' inside the DTP or SVC package logs.

    The cause of the error is a loss of the network connection to the detector. The error on the DCC will be this:

    In order to fix this issue the detector computer must be reset following the Old Full Shutdown Procedures and the Old Full Startup Procedures for the offending detector.

    The volumes will need to be remounted and the instrument control software should be reset.

     

     

    In the case that a network cable will be disconnected while the detector computer is running follow this procedure:

    To prevent this kind of situation, we have to do following works before detaching network cables. 
    0) Close the all components of IGRINS Control Software and eject 
         all mounted volumes. -> [Turn ICC off] is the simplest way.
     
    1) Stop socket servers on DCCs   
         There is the 'Enable Socket Server' option in the 'System' menu on the 
         Teledyne's socket server. We can stop the socket server 
         by uncheck the 'Enable Socket Server'.
    2) Turn Lakeshore350 off
         To avoid malfunction of the embedded socket server of LS350.
     
     After re-connecting network, restart the Teledyne's socket servers 
     (check the 'Enable Socket Server') , turn LS350 on agin
    , multiple selections available,

    Confluence Documentation | Web Privacy Policy | Web Accessibility

    University Wiki Service

    IGRINS
    • File Lists
      File Lists
       This trigger is hidden
    Results will update as you type.
    • People
    • IGRINS Newsletters
    • Publications using IGRINS
    • Downloadable Software
    • IGRINS at McDonald Observatory
      • Current Manuals - McDonald Observatory
      • Old Manuals - McDonald
        • Old Observer Cookbook - McDonald 2.7m
        • 2014 Troubleshooting
          • Troubleshooting Guide
          • How to reboot Colossus if TCS fails
          • Solution to Slow Detector Display on ICC
          • Detectors not taking exposures
          • Files not saving at end of exposure
          • K band Connection is Dead
          • FAQ
        • Old User and Proposer Information
        • Old Observing Procedures
        • Old Instrument Team Procedures
    • IGRINS at the DCT
    • IGRINS at Gemini South
    • IGRINS Summary Documents
      Calendars

    You‘re viewing this with anonymous access, so some content might be blocked.
    {"serverDuration": 11, "requestCorrelationId": "291df784a5b64ac3a0223a746dd9d259"}