Plasmatherm #2
Nick's Notes:
David's Notes:
Left chamber is for etching Si and Right is for etching SiO2 and Si3N4
Chambers are left in vacuum. Must vent to load
utilities → select active → (select right/left)
**if vent is greyed out, press standby on bottom left, then go back to select vent**
**if this still doesn't work, press on, then standby**
Cleaning
wipe down stage and showerhead with water or IPA on a clean wipe
only wipe showerhead in one direction.
Close chamber then pump down to low vac
utilities → pump chamber → low vacuum
O2 clean - should always perform O2 clean before using the tool and 5 min O2 clean after 45 min of use
process → load
select "o2 clean.prc" recipe in main recipe folder
Running process
Find recipe in
**if any problems arise, click "end step" until out of problem step, not abort **