Transcardiac Perfusion

Transcardiac Perfusion


 

Safety Precautions

  • You must complete required lab safety training before starting this procedure.

  • You must complete required animal use training and listed on IACUC protocol before starting this procedure.

  • If this is your first time doing this procedure, ask to be trained by an experienced lab member.  If you have not done this in a while, you should ask for a refresher.

  • Also review the following material:

    • Kuwajima M, Mendenhall JM, Harris KM (2013) Large-Volume Reconstruction of Brain Tissue from High-Resolution Serial Section Images Acquired by SEM-Based Scanning Transmission Electron Microscopy. Methods Mol Biol (Nanoimaging: Methods and Protocols) 950:253-73. (PDF)

  • Further reading:

    • Hayat MA (1981) Fixation for electron microscopy. Academic Press

    • Karnovsky MJ (1965) A formaldehyde-glutaraldehyde fixative of high osmolality for use in electron microscopy. J Cell Biol 27:137A-138A. (PDF)

    • Tao-Cheng JH et al. (2007) Structural changes at synapses after delayed perfusion fixation in different regions of the mouse brain. J Comp Neurol 501:731-740. ()

    • Dehghani A et al. (2018) Nuclear expansion and pore opening are instant signs of neuronal hypoxia and can identify poorly fixed brains. Sci Rep 8:14770. ()

  • Before starting, even if you have done this procedure before,

    • read this protocol entirely.

    • If you are using adult rats, review the instruction movie on the tracheotomy procedure.

    • review relevant Safety Data Sheets and Harris Lab SOP (also see below).

    • ensure you have all reagents and supplies listed below.

    • ensure all equipment is in good working order.

      • Isoflurane vaporizer must be serviced annually; check tubing, surgical instruments, etc.

    • have all waste containers ready (also see Clean-up).

    • plan your schedule well so that you wouldn’t be rushed to complete the procedure.

  • Review SDS and Harris Lab SOP for the following hazardous chemicals used in this procedure:

    • Ethanol: flammable; irritant (eye)

    • Formaldehyde

    • Glutaraldehyde

    • Hydrochloric acid

    • Isoflurane

    • Sodium cacodylate: carcinogen; irritant (skin, eye); skin permeator

    • Sodium hydroxide

  • Other hazards associated with this procedure include:

    • Physical: compressed gas cylinder (O2/CO2)

    • Sharps: Surgical scissors, scalpel, needles

    • Biological: Live rats/mice, blood, animal waste and soiled cage bedding

  • The following Personal Protective Equipment is required for this procedure:

    • Lab coat

    • Nitrile gloves (double-layer required; regularly check for holes)

    • Eye goggles

    • Mask

    • Face shield

  • Place a piece of absorbent sheet on the work surface before starting the procedure.  When done, discard into the “Solid Waste – No UA” bag


 

Reagents, Supplies, Equipment

Personnel Protective Equipment

  • Eye goggles, face shield

  • Gloves (nitrile)

  • Lab coat

  • Mask

  • Plastic Apron (optional)

Assorted tubing, connectors, and clamps

Air Pressure System

  • Sphygmomanometer with a hand pump (cut off the cuff)

    • Use hand pump in the absence of an air pump.

  • Air pump: Perfusion Two (Leica Biosystems 39471010)

    • This pump is designed to deliver constant air pressure at 300 mmHg. To be able to adjust pressure, we added a regulator (Air Logic R-7010-2B2B-W/K) between "Air Tank" and "Perfusion Pressure" switches (See Figures 2-4).

  • Air filter (Parker Hannifin 9933-05-AAQ)

  • 1-L glass bottle (Kimax, or equivalent) × 1

  • #6 two-hole stopper × 1

Oxygenation System

  • O2/CO2(95%/5%) cylinder with regulator and support

Anesthesia System

  • Isoflurane (Animal Health International 19632158; stored at RT in well-ventilated area)

  • Medical grade O2 cylinder with regulator and support (Can be replaced by O2/CO2 or an oxygen concentrator)

  • Large glass desiccator, wad of 4-5 Kimwipes, Pasteur pipet

  • Matrx VIP3000 Isoflurane Vaporizer (Stoelting)

    • This apparatus must be serviced annually.

    • For Central Texas area: William Connery, Handlebar Anesthesia, (512) 423-7668, handlebaranesthesia@yahoo.com

  • A small animal ventilator (Harvard Apparatus Model 683, or equivalent)

  • Nose cone (e.g., Stoelting; Kent Scientific)

  • endotracheal tube (a modified 16-gauge hypodermic needle) with Y-connector

  • Optional: Plexiglas anesthesia induction chamber (e.g., Stoelting; Kent Scientific)

Perfusate System

  • Deep water bath (e.g., VWR Digital Unstirred Water Bath, L×W×D = 127/8"×1113/16"×529/32"; VWR 89032-216)

  • Lead donuts × 2

  • #6 stoppers: three-hole × 1, two hole × 1

  • 3-way valve × 1

  • Flow regulator × 1

  • 13-gauge needle with 60° bevel with cork disk ½ inch from tip (to restrict depth of penetration; use a16-gauge for juvenile rats)

  • Air stone × 1

Waste Collection System

  • 4-L vacuum flask (preferably plastic-coated) × 1

  • #11½ one-hole stopper × 1 

  • Vacuum line filter (Whatman VACU-GUARD, VWR 28137-858), 1 per day of perfusions

Dissection/surgical instruments

Scissors and knife etc. should be sharpened periodically.

  • Stainless steel dissection tray with Styrofoam board cut to fit snugly lengthwise in the tray with ~1-inch space on one side. 

  • Pins to secure animal to Styrofoam board

  • Scalpel and blades

  • Assorted dissecting scissors

  • Forceps

  • Hemostats

  • Retractors

  • Knife (to decapitate rat)

  • Bone Rongers

  • Spatulas

  • Silk suture (or cotton tread from fabric store can be used instead)

  • Gauze pads

  • Applicator sticks

 

For Preparation of Perfusates

All reagents are stored at RT unless otherwise noted.

  • Calcium chloride dihydrate (CaCl2·2H2O; Sigma-Aldrich 223506)

  • d-Glucose (Sigma-Aldrich G7528)

  • Formaldehyde (20% aqueous solution in 100 ml bottles; Ladd Research 20304)

  • Glutaraldehyde (50% aquieous solution in 100 ml bottles; Ladd Research 20211; stored at 4°C)

  • Hydrochloric acid (HCl; 1 M aqueous solution; for adjusting pH)

  • Magnesium sulfate heptahydrate (MgSO4·7H2O; Sigma-Aldrich M5921)

  • Potassium chloride (KCl; Sigma-Aldrich P9333)

  • Sodium bicarbonate (NaHCO3; Sigma-Aldrich S6297)

  • Sodium cacodylate trihydrate (Ladd Research 20305)

  • Sodium carbonate (Na2CO3; Sigma-Aldrich 223530)

  • Sodium chloride (NaCl; Sigma-Aldrich S7653)

  • Sodium hydroxide (NaOH; 1 M aqueous solution; for adjusting pH)

  • Purified water

    • double-distilled, ASTM type I, WFI (water for injection), or equivalent; e.g., Fisher 91-502-5

    • ASTM type I water can be made with a properly maintained water filtration system (e.g., Milli-Q® Type 1 Ultrapure Water System )

  • 2- or 4-L glass beaker × 2

  • Magnetic stirrer × 2

  • 100-ml glass beaker and a disposable pipet (for dissolving sodium carbonate [Na2CO3])

  • 1- or 2-L graduated cylinder (depending on the final volume)

  • pH meter (and calibration standards)

  • 1- or 2-L glass bottle (Kimax, or equivalent) × 2

  • Vacuum filtration system (pore size = 0.22 µm; e.g., VWR 10040-468 – this one screws onto Kimax glass bottles)

  • WESCOR VAPRO 5520 Osmometer (with calibration standards, filter paper disks, pipet, and pipet tips)

Record Keeping

  • Timer (in count-up mode)

  • Worksheet