WESCOR VAPRO 5520 Osmometer
BEFORE USE:
- Keep the instrument on.
- Make sure the instrument is in the standby mode (displaying "READY >").
CALIBRATING THE INSTRUMENT:
- Choose the correct standard for your sample
- 290 mmol/kg for ACSF and KRC buffer.
- 1000 mmol/kg for fixatives
- Open a fresh ampoule of the standard.
- Turn the sample chamber lever on the right side to the "OPEN" position.
- Use the "SELECT" keypad to select "AVERAGE MODE", then press "ENTER".
- Pull out the sample holder.
- Place a single sample disc into the sample holder (shiny metal well) with the supplied forceps. NEVER touch the metal well with the forceps.
- Pipet the standard onto the sample disc with the pipettor tip resting in the notch of the pipettor guide. Do not touch the sample holder surface with the pipettor.
- Push the sample holder back in, and turn the sample chamber lever to the "CLOSE" position. If this is the third measurement, skip to Step 12.
- The display will indicate "IN PROCESS", and the chime will sound when it's done. The osmolality of the standard is displayed at this point.
- Open the sample chamber, pull out the sample holder, and remove the specimen from the holder immediately after a measurement. Use Kimwipe to carefully remove the wet disc and any residual liquid from the holder (NEVER use forceps). If necessary, clean the holder with Kimwipe moistened with ddH2O.
- Repeat Steps 5 - 9 once more, and then Steps 5 - 8.
- At end of the third measurement (keep the sample chamber closed), press "SELECT" for "CALIBRATE", then press "ENTER".
- Clean the sample holder as described in Step 10.
- Push the sample holder back in, and turn the sample chamber lever to the "CLOSE" position. You can measure your sample once the instrument display “READY>”.
MEASURING THE OSMOLALITY OF YOUR SAMPLE:
- Turn the sample chamber lever on the right side to the "OPEN" position.
- Use the "SELECT" keypad to select "AVERAGE MODE", then press "ENTER".
- Pull out the sample holder.
- Place a single sample disc into the sample holder (shiny metal well) with the supplied forceps. NEVER touch the metal well with the forceps.
- Pipet your sample onto the sample disc with the pipettor tip resting in the notch of the pipettor guide. Do not touch the sample holder surface with the pipettor.
- Push the sample holder back in, and turn the sample chamber lever to the "CLOSE" position.
- The display will indicate "IN PROCESS", and the chime will sound when it's done. The osmolality of the sample is displayed at this point.
- Open the sample chamber, pull out the sample holder, and remove the specimen from the holder immediately after a measurement. Use Kimwipe to carefully remove the wet disc and any residual liquid from the holder (NEVER use forceps). If necessary, clean the holder with Kimwipe moistened with ddH2O.
- Repeat Steps 5 - 9 once more, and then Steps 5 - 8.
- At the end of the third measurement (before opening the sample chamber), average of the three measurements are displayed. Record this in your notebook.
- Clean the sample holder as described in Step 8.
- Push the sample holder back in, and turn the sample chamber lever to the "CLOSE" position.