Wafer Thinning
Wafer Thinning
Turn hot plate on to level 4 (corresponds to ~130C)
Crystal bond melts at 121C
Measure the starting thickness of the sample
Zero out the thickness measuring tool by dropping the arm and pressing “zero”
Place the sample under the arm, drop it to measure the starting thickness
To replace the battery, twist the face and pull off to pop out watch battery
Get the “step 1” holder
Only “step 1” is needed to sufficiently thin the wafer for a good cleave
Steps 2-3 are for wafer polishing (might be necessary for reflection/transmission measurements)
Take out the glass chuck from the “step 1” holder, stick the “step 1” silicon carbide paper onto it
Replace the glass chuck into the holder and fit the holder into the thinning machine
Place the sample holder (stored in Briggs’ box) onto the hot plate and let it heat up
Use the razer to chip off some chunks of crystal bond and use the crystal bond tools to place a pea/dime sized amount of the chunks onto the sample holder
When the crystal bond is melted, place the sample epi (or glass slide) down (substrate up)
Rotate sample a bit to even out the crystal bond
Don’t let it sit on the hot plate too long or the resist can get over-baked
Remove the sample holder (use 2 pairs of tweezers) and let it cool for 5 minutes
Spray DI water on the sandpaper (quarter-sized amount)
Place the sample substrate down and connect the sample holder to the thinning tool arm
Press go (covered in slurry) and let run for 30 minutes
When complete, lift up and remove sample holder
Drop sample holder on hot plate and remove sample using crystal bond tools
Let crystal bond solidify before placing sample down anywhere
Take out holder and peel off sandpaper
Wipe with acetone
Remove wafer bond from sample holder with acetone
Clean off with acetone in acetone dish
Dump dish down “sink” (ensure waste jug not full)
Wipe down
Measure thickness of sample at this point
Repeat steps 3-17 as needed
Turn everything off and put away
Attaching to a Glass Slide
Cleave off a square of a glass slide in the bat cave
Heat up hot plate to 90C
Place the sample at the resist spinner an apply thick resist at 2000 rpm for 40 sec
Splat the sample epi down onto the center of the glass slide (substrate exposed)
Place the glass slide down onto a hot plate at 90C for 1 minute
Check to make sure it is indeed bonded together
Removing the Glass Slide and Gunk
Fill the PG remover dish with PG remover
Place sample + glass slide combo in the dish and gently (GENTLY) slide the sample off
Use 1 pair of tweezers to push the sample parallel to the glass slide and use another pair to push against the glass slide
Fill the acetone dish with acetone
Move the now-free sample to the acetone dish and let swim in acetone to remove the rest of the resist and any residual III-V gunk
If lots to remove or going slow, make a little foil hat and let soak in acetone overnight
Can also sonicate if it still won’t budge (carefully!!!)
Toss the glass slide in arsenic trash
Fill the IPA dish with IPA
Soak the sample in IPA to finish cleaning