Laser Testing Procedure
Laser Testing Procedure
Written from NTS notes NTS, 9/15/2015
Note: Must have a laser sample processed and ready to test.
Prep
Setup detector
Fill dewar with LN2 (from loading dock)
Put InSb detector on spectrometer, be careful with it
Fill detector red reservoir with LN2, in portions
Pour til LN2 spouts out (this is cooling the reservoir), pause
We know it’s cold when the fog is tiny.
Pour LN2 in to fill reservoir
Don’t pour too fast
Watch for detector overflow and stop pouring
Remove funnel and wait for LN2 geyser to stop; fog will now just escape from hole. Top off.
We know it’s full when overflowing after a top off or two.
Do not turn on detector til it’s at 77K
Do first! InSb detector needs LN2 cooling. Do not turn on detector without LN2. Always turn detector off.
Mount sample with screw-bolts
Setup Laser Test jig
Slide jig into place (almost)
Extend PL setup rails
Put on lense, slide jig into place.
Turn on green laser (slowly) and align jig. Only need 0.9-1.3 amps.
Shutter laser with iris when done (for now)
Setup vacuum for probes
Connect red tubes on probes to vacuum line (clear green)
Seal connection with kapton tape
Turn on vacuum
Make System Connections
Probe cable to T junction on vector board
Back of TE cooler to other BNC
The vector board connects to a PCB resistor tuning network
Connect PCB current output to Oscope
Cap other BNC on PCB with 50 Ohm cap
Change lock-in ref-in cable to back of pulsed current source (trigger out T)
Connect chuck BNC cables
Connect chiller hoses
Connect TE cooler cable to stage
Turn on most electronics
Turn on Oscope
Turn on alignment laser driver (align laser sample soon)
Turn on chiller (box under table)
Flip both switches on chiller box
Check for leaks where hose meets chuck
Turn on TEC control/laser driver
Set TEC temp and turn on output
Can switch from TEC to current with knobs
Position system components
Align system, lens in front of sample
Position microscope
Get probes in place, do not land!
Do not land probes til current supply is on
Put laser mount on chuck and screw in to front screwset (by chuck edge)
7a. Turn on lock-in amp
Turn on the current supply (output should be and stays off)
Supply output will be off
Land probes carefully
Unshutter green laser to perform alignment
Want laser as dim as possible while still visible to align
Do coarse alignment (no microscope) first by adjusting lense (on rails)
Want to get as small and bright a spot as possible, with smaller and a little dimmer better; we’re going for ballpark here.
Now use microscope and do fine alignment with micrometers.
Get green hazy spot at right level and XY alignment to be in line with laser bar.
Turn on spectrometer
Setup spectrometer software (was with old labview software)
Setup your grating and filter wheel
Filter wheel is needed to choose lambda range (and avoid signal from substrate PL)
Turn on current output
Set pulse width to 0.2 us and rep rate to 100 us (10 kHz)
Turn up current (~50-100 mA/s) [seems fast]
Wait for lock-in amp to show signal
Typically channel 1 will go from bouncy # to large # ( >8 uV) suddenly, e.g. after one amp input
Test
Setup and run spectrometer scan
Setup PL software [for old labview]
Choose filter (150 4) and slits
Set start position and hit Go To to setup spectrometer
Look at the center of this spectrum
On current box press display to toggle between current and pulse width
Change lock-in to positive edge triggered if needed
Set PL software endpoint and step, and check start position (press Go To just to be sure)
Setup filepath, sample ID, notes, etc.
When you increase current, you’ll get electroluminescence
Start Scan
E.g. Today we’re starting at 1500 mA and increasing by 200 mA each run
We’re starting each line at 3200-3700 nm and then focusing in as needed
Turn current output down and off (and back on) as needed
Look at the center of this spectrum
To finish:
Take current down and turn off output
Lift up probe
Move microscope out of way and turn off
Turn off chiller and TE cooler
Turn off detector and remove sample
Turn off green laser