Slit View Camera Package not taking images

Slit View Camera Package not taking images

First try the following:

If you notice the Slit View Camera (SVC) package not taking images (e.g. there is a yellow “SLOW” and a red “FAILURE” notice):

  • Stop the guiding.

  • Abort the current Sequence using the DTP.

  • Try taking a single with the SVC. If the slit camera recovers and reads out the next image successfully, resume guiding and data acquisition. Otherwise continue with the troubleshooting steps.

  • Close the SVC Package. This might take a minute or two with the mouse turning into a spinning beachball and the screen reducing in size. Once the SCV GUI is closed, check the ICS Hub to ensure the “run SCP” shows “disconnected” otherwise click “Check Processes” on the ICS Hub and click the “Kill” button next to the active SCV GUI option.

  • Go to the IGRINS Hub and reload the SVC package by clicking the “run SCP” button.

  • Take a single exposure to make sure it works. If not, try resetting the SVC package again. If that fails, you might need to restart the S camera entirely (see Detector Power Cycling and Re-initialization) and, if that fails, restart the software on the SVC detector control computer (see the full startup procedures, if you don’t know how to do this contact an IGRINS Team member).

If troubleshooting is successful:

  • Remember to update the various options in the SVC package to what they were before the failure.

  • Start guiding on your target again. Make sure the guiding is good before starting your spectra acquisition.

  • Re-input the target information in the Data Taking Package since resetting the SVC package will erase whatever was there previously.

    • Click From TCS on the DTP, re-enter your target name, re-enter the object type, and then click apply. Double check your exposure time and sequence are what is expected.

  • If everything looks good, resume data taking.