FTIR
This page discusses the FTIR system.
Last Updated by: AFB 20160920
Introduction
Fourier Transform Infrared Spectroscopy (FTIR) is a measurement technique that quickly analyzes the optical transmission and reflection properties of a sample.
LASE FTIR System
Bruker Vertex 80v FTIR with a Bruker Hyperion Microscope
(1) Infrared mode | (2) Visible mode | (3) Reflection mode | (4) Transmission mode | (5) Temperature indicator for LN_2 cooled detector |
(6) (Obstructed) Aperture wheel with pinhole | (7) Condenser height adjustment knobs | (8) (Obstructed) Condenser set screws for XY adjustment | (9) 36x IR microscope objective | (10) 4x microscope objective |
(11) Controls for objective aperture. Allows for precise area of measurement at the cost of signal | (12) Rod that controls direction of light either to microscope eyepiece or to camera | (13) Front LCD screen power |
Manuals
Hard copies of the manuals are located in the FTIR lab. We are still trying to procure digital copies.
Software
Bruker provides free OPUS software for off-site data viewing and analysis. A download to that software is here:
OPUS viewing software download
The software will download as a .zip folder. Extract the contents of the .zip to a directory of choice on your computer. Then run OPUS.exe (if will show up as the OPUS icon in the folder).
To use the viewing software, the output data/spectra must be saved in an appropriate file format. Once a .o file is loaded into OPUS, use save as, then select the Mode tab to pick the file format. The following file formats should work for both using the viewing software and for any data manipulation you might need to do:
Data point table (.dpt)
JCAMP DX
Galactic
The .dpt is the recommended option, because that is what everyone else in our group uses.
NOTE: OPUS files, i.e. .o files, cannot be viewed with the viewing software
Training
Get trained on the FTIR by one of the senior students. Be sure to read through the procedure below before starting a training session
Procedures
Set-up
Add LN2 to the top of the microscope
Try not to overflow the LN2, because it wears down the enamel
Fill until the LN2 holding tank is full
Turn on nitrogen purge line using valve by the back of the microscope
Start Opus software
Password: OPUS
The microscope is at cryo temperatures and ready for use when the green light (bottom right corner of the screen) is on. It is too warm and not ready for use if the light is red
Turn on the microscope (black switch along the back of the microscope, along the bottom and toward the backroom door)
IF NEEDED: Change the beamsplitter for the range of measurement desired
Vent optics (~ 2min)
Go to Measure/advanced measure/Basic and select "Evacuate Optics"
^ same for purging optics
Add correct beam splitter
Update trolley sheet Reinsert the beamsplitter, using red tab to secure the beamsplitter
vent system (~2 min)
Advanced Data Collection Transmission Measurements
This measurement technique allows for greater control of data collection and is faster.
There are two options for stage control in OPUS
Open "motor stage control" and check "activate joystick"
Open “microscope wizard” and select Hyperion 2000 option
Calibrate transmission measurement
Switch microscope to visible and transmission mode
Close microscope aperture (6) to pinhole
Adjust focus of beam of light to a circle by raising and lowering condenser (7)
This can be viewed through the microscope eyepiece or the monitor with video wizard depending on the position of the microscope rod (12)
Optimally it should look like the figure to the right
If the condenser isn't in the cross hairs, adjust its position using the setscrews on the back (8)
If grossly misaligned the microscope eyepiece will give a large field of view to locate the condenser
Open aperture
Switch microscope to infrared mode
Check counts by going to advanced measurement settings>check signal>ADC Count
Counts for the global background should be between 10,000-14,000
Taking global background
Click "advanced data collection"
Choose file path in "advanced" tab
Have "single channel" as the only data block to be saved
On the "basic" tab choose your sample description
Let purge for 1-2 minutes
Sample single channel
Sample measurement
Load sample, make sure that the area of interest is over the hole in the stage
Switch to visible and reflection mode
Find area of interest with the microscope
Switch to infrared and transmission mode
Maximize counts by moving the condenser and the sample stage with joystick
Take measurements
Microscope Wizard Reflection Measurements
The general procedure for the reflection measurements is on P.27 of the Hyperion user manual
Complete the Set-up above
In Opus, select "Start Video Wizard"
select Hyperion 2000
"start"
Move the condenser out of the way, i.e. all the way down so any x-y stage movement does not catch the top of the condenser and strain the stage motors
OPTIONAL: Reset the xyz stage
Only need to do this if the stage was altered significantly
Manually change the objective
On the microscope, move to the 36 X IR (large black objective w/ gold band that says "(36X) NA=0.5")
In Opus, update the objective option in the GUI to 36 X IR
Add sample to the stage
Find and focus on the sample with the microscope by using the joystick
Set to visible mode. Note that the 36 x IR objective will still work for visible light
Find the particular sample features that you want to measure
Recommendation: focus on the edge of the sample or some feature edge, then move laterally (x-y) to the desired position on the sample
Capture image in a region of the sample where the measurement will take place.
Take background measurement
Move sample out of the way with joystick and put gold mirror from blue box in the way to collect a background signal for reflection
In visible mode, adjust the focus height to get a focused position on the mirror. Again, focus on the edge first before moving somewhere nice-ish on the mirror surface (there are many crack/scratch marks)
Change to IR (squiggly waveform)
Check that the polarizer is in the IR mode (i.e. almost entirely pushed in)
There is a sliding knob on the left side of the microscope
In Opus Measure/Check signal
Change to ADC mode
Adjust the height using the joystick (twisting for z control) to maximize the signal counts
Things to keep in mind:
Make sure the IR polarizer, 36xIR, and IR light source settings are being used!
Get to a rough max intensity for the IR counts by manually adjusting the joystick, then use "Auto Align" followed by "Fine Auto Align"
Typical gold mirror peak ADC count value is ~4450
Measure the background once, i.e. only collect one background measurement point
Take sample reflection measurement
Move back to the sample position using "move xy stage" and click in the far right window
Note: this only moves in x and y. You will still have to adjust z height, which is described below
Change to visible
Readjust the height/focus back to the sample
z height difference between a glass slide and the gold reflecting mirror is ~2500 um
z height difference between a Si wafer (350 um thick) and the gold reflecting mirror is ~2700 um
Change to IR, IR polarizer, and 36 x IR
In Opus Measure/Check signal
Change to ADC mode
Adjust the height using the joystick (twisting for z control) to maximize the signal counts
After rough adjustment with joystick, use "Auto Align" followed by "Fine Auto Align"
Some typical ADC max count values:
Bare Si (350 um thick) = 1450
Set measurement position(s)
Pick a set of data points for where on the sample to collect data. There are several option in Opus, e.g. grid.
Recommendation for homogeneous sample: use 5 arbitrary/random points on top of each other. In theory the output data should be 5 ~identical columns
Set data file path information
Sample name
File name (use the sample name)
Set directory path (desktop/user data/<your file>)
Run scan (~2-3 min)
Do not repeat the scan (unless you need to). Select "End"
Viewing and saving the data of interest
Select the data block of interest, for reflection it is the RFTL block
Save as data point table (.dpt) format, which acts as a .csv file
Microscope Wizard Transmission Measurements
The general procedure for the transmission measurements is on P.30 of the Hyperion user manual
Complete the Set-up above
In Opus, select "Start Video Wizard"
select Hyperion 2000
"start"
Move the condenser out of the way, i.e. all the way down so any x-y stage movement does not catch the top of the condenser and strain the stage motors
OPTIONAL: Reset the xyz stage
Only need to do this if the stage was altered significantly
Manually change the objective
On the microscope, move to the 15 x IR objective
In Opus, update the objective option in the GUI to 15 x IR
Add sample to the stage
Find and focus on the sample with the microscope by using the joystick
Perform in reflection mode
Set to visible mode. Note that the 15 x IR objective will still work for visible light
Find the particular sample features that you want to measure
Recommendation: focus on the edge of the sample or some feature edge, then move laterally (x-y) to the desired position on the sample
Capture image in a region of the sample where the measurement will take place.
Take background measurement
Change to transmission mode
Select "Measure background once" and "User defined position"
Change to visible and transmission mode
Move the sample out of the way such that all the light from below clears the edge of the sample and the edge of the stage hole
Optional: adjust the condenser with the coarse alignment knobs until you see a washed out bright white light
In visible
Adjusting the condenser silver knobs until the until the video is washed out with bright light
Close the pin hole via the black knob behind,below,back-left of the stage
Adjust the silver knobs to bring the light into focus so it looks like a circle
Adjust the the silver set screws at the base of the condenser to position the center beam/circle in the crosshairs of OPUS
Open the pin hole
Switch back to IR
ADC counts should be ~18000
Wait 2-5 minutes to allow the nitrogen to purge the atmosphere in the measurement box
Collect background
Align sample
Move x-y stage, click in the far right window to actually move the stage to the sample position
Set the measurement points (grid, random points, etc.)
Change to IR mode and go to Measure/Check Signal
Adjust the condenser height to maximize the counts
Adjust the stage via joystick to maximize the counts
Again adjust the condenser height to maximize the counts
Collect sample transmission data
Set data file path information
Sample name
File name (use the sample name)
Set directory path (desktop/user data/<your file>)
Run scan (~2-3 min)
Do not repeat the scan (unless you need to). Select "End"
Viewing and saving the data of interest
Select the data block of interest
Save as data point table (.dpt) format, which acts as a .csv file
Turn off system
After finishing a measurements:
Close the Opus software
Turn off the nitrogen feed valve
Turn off microscope using switch on the back of the microscope
Troubleshooting
The Opus software frequently crashes. Restart the software when this happens. Note: the entire measurement must be restarted, including the background
If you need to power cycle the spectrometer
Wait until all the lights are green on the back/top/left side
After restarting Opus, you must purge the optics in Measure/Basic Purge Optics