Bravo MBE Junk Growth
Note: Always check CAR and cracker motors for PID vibrations.
What is the Junk growth, and why do it?
The junk growth is the first growth performed when bringing up a system. The philosophy is after opening up the system for maintenance (i.e. sources are cool and the chamber is at atmosphere), there is a a lot of "junk" that has landed and built up on, among other things, the source material, cells, and cell shutters. The junk growth is a chance to heat all of these up past their typical operating temperature to blow off all junk and to test whether the cells are operating within an acceptable range.
The growth itself has the wafer (full 3" Si (100)) facing the sources. The shutters on the sources are left open while the cells heat up. Once the cells reach the junk growth temperatures, a series of shutter cycling occur. Cycling heats up and cleans the backside of the shutters using the cell's heat. A group V needs to always be on the shutter during growth; otherwise, the sample becomes drippy with molten group 3 material.
Substrate rotation is not used because the orientation of the wafer needs to be preserved. Problems with the shutters and sources can be seen at the different regions of the wafer -- this can be a potential troubleshooting tool is something goes abnormal happens during the growth.
In addition to monitoring AMBER and the rest of the electronics during the growth, also watch the actual shutters and listen to them when actuating during the cycling portion of the recipe. This is a good time to figure out if there are any shutter problems.
Day before junk growth
Load 2 full 3" Si (100) wafers and 4 1/4" SI GaAs wafers onto the trolley. The Si wafers will be used for the pre- and post-melting of aluminum. The 1/4 GaAs pieces with be used for the Std. quantum well and superlattice (SL) growths both pre- and post-melting of aluminum
Work with an older student if you are unfamiliar with the loading process
The 3" Si wafers need another 3" Si baking wafer
The 1/4 SI GaAs pieces need sapphire backing wafers
After loading samples back into the LC, perform a LC bake (standard recipe on growth computer in C:\AMBER\recipes)
There is no STO (source turn on) for the junk growth
Check your recipe with the older students. Sometimes junk growths temperatures and the number of shutter cycles change, e.g. replacing a shutter likely will require more shutter cycling to check its performance.
The junk growth takes most of the day to complete. You need to stay in the lab (or have another student monitoring the machine) during the shuttering section. You will not need to unload the trolley at the end of the day (if there are no problems).
Growth Day
Recipes
Below are examples of previous junk growth recipes. Check with older students for the most recent and correct recipe to run as the junk growth parameters change over time, based on the status of the machine...
Junk growth without aluminum
Junk growth with aluminum
Not actually a junk growth, do an outgassing recipe
If there's a power failure but Al stays melted, you need to do another junk growth with Al (rather than outgassing)
Bake Si Wafer on heater station
Move Trolley into buffer chamber (after LC lamp bake)
Move Si wafer onto heater station (HTS)
Move trolley to Echo buffer or to the far side of the Bravo BC (i.e. closest to the LC gate valve). The Si wafer is unclean and will outgas crud back onto the trolley if the trolley is not moved.
End and Save RGA scans on GC and BC, if applicable. Ask the older students before stopping the RGA scans, because they may be monitoring them.
Save scans (use the YYYY-MM-DD convention followed by a brief description of the scan)
Turn off electron multiplier then filaments
Disconnect the COM in the software
Turn off (i.e. flip switch) on the actual RGA head/box
If needed, unplug the serial cable to the RGA head/box then unplug the power cord
Follow the steps above in reverse for turning the RGA back on
Leave chamber ion gauges (IGs) on
In the AMBER config menu, adjust the max temp set point of DD Be from 400 --> 850
We de not intend to use the DD Be cell for anything, so we intentionally keep the max temp low
To properly clean out the cells, we change the max temp before starting the junk growth, because you cannot adjust AMBER config settings while a recipe is running
Once the recipe is finished at the end of the growth day, we need to change the AMBER DD Be config max temp back from 850 --> 400
Compile and run the junk growth recipe
Recipe has heater station (HTS) bake, waitop command, cell temperature and ramp rates
HTS bake is 450 C for 1 hour
Make sure all viewport lights on the BC are turned off when starting a HTS bake
Prep for Growth
Wait for HTS to drop below 250 C before transferring wafer of the HTS and back onto the trolley
Turn off the Beam Flux IG
Transfer wafer to GC (CAR) after HTS bake
CAR must be in 'local' and the 'transfer' position. The CAR rotation must be in 'local' and the rotation turned off
Set CAR temp Eurotherm to 'manual' with the output power at ~5-7% for transfer
Open GC to BC after HTS is under 200 C to prevent outgassing into the GC from the BC
Open GC gate valve (GV) first, then GC shutter (fully open)
Reverse order to close after loading
Always close the shutter!
Remove RHEED block first and put it onto the trolley.
Transfer wafer in
Set CAR temp Eurotherm to 'auto', which will bring the substrate temp back to 200 C
Rotate CAR to growth position
Flip switch to CAR position, should be at 'transfer' or ~352 degrees
Put in 'manual' (from 'local') and rotate it (by hand) down to under 5
Do not tighten the set screw on the CAR rotation knob! Leave it loose
Put into 'local' and hit 'growth' to send it to the growth position or 0 degrees
There is no away bake for Junk Growth
The CAR temperature is 200 C (idle) during the entire junk growth
You don't want to heat the wafer and redeposit the junk all over the rest of the system
Leave the HVP on at the idle value of 105 Watts
There is NO rotation for junk growths (so we can see if there are any issues with cells and shutters that can be debugged with nonuniformities on the wafer)
Start the RGA scan if needed. Leave the shutter closed
Make sure all GC viewports shutters are closed
Junk Growth
Continue the recipe by finishing the 'Starting junk growth' waitop command
The recipe will heat up the cells with shutters open
Check sublimator power/temp mode, and double check cell set points and ramp rates
Adjust the trend charts so you can see all the cell temps at once
NOTE: just because the cell temperature has flatlined, it may be at the wrong temp!!!
Check that the cells are reaching the desired temp
There may be max temp or power setting that needs to be adjusted
After the cells heat up, there is a waitop command to re-home the cracker valves
Reopen the Cracker valves to 250 before continuing the waitop
Note: Don’t close Group V cell shutters when the Group V cell needle valve is open.
Keeping a Group V overpressure is necessary to prevent an accumulation of molten Group III on the wafer.
The recipe will keep either As or Sb open at all times
Recipe will cycle cell shutters, closes effusion cells, and run source turn off
Listen to and watch the shutters (in AMBER, on the switch panel, and the actual shutter rods) to identify any malfunctioning shutters
After the source turn off starts, there is a waitop command to re-home the cracker valves
'Remote' --> 'Local'
Hold down 'Local' and 'Home' simultaneously
You will be prompted with 'Do you want to do manual calibration?' --> 'No'
'Test full travel' --> 'No'
Set back to 'Remote'
After re-homing the cracker valves and hitting continue on the recipe, the As, Sb shutters will close
Finish procedure by removing Junk growth wafer from GC
Change the max temp set point of DD Be back to its lower value, i.e. change from 850 --> 400 in the AMBER config menu
Prepare system for the standard QW and SL growths. Assuming you also loaded the 1/4" GaAs pieces for the QW and SL growths, you do not need to unload/reload any wafers at the end of the current growth day
Growth Parameters
From Wiki page System_Bravo_Junk_Growth_Temps
NOTE: the temps below change over time!!!!
We adjust a temp up if more of a particular source is needed. See the last junk growth recipe and discuss what temp is best with the older students.
As described in the "What is junk growth and why do it?" section, we heat the cells past their warmest operating temperatures during growth. Determine what the hottest temp is during growth for each cell by talking to all MBE growers in the group, then increase the junk growth temp by ~50-100 C.
Adjusting the max temp might require you to adjust the max temp setpoints on the system (Sorensens, Eurotherms, AMBER). Ask an older student how to check these if you decide to increase the max temp from the last junk growth.
Eurotherm | Temperature |
|---|---|
As Sub | 415C |
As Cracker | 900C |
Sb Sub | 65% (~536C) |
Sb Cracker | 950C |
In Tip | 950C (depends on InTip growth temp and InTip-InBase temp) |
In Base | Let AMBER do it |
Ga Tip | 1070C (depends on GaTip growth temp and GaTip-GaBase temp) |
Ga Base | Let AMBER do it |
Bi Tip | 685C (depends on BiTip growth temp) |
Bi Base | Let AMBER do it |
Er | 1020C (depends on ErTip growth temp) |
Si | 1350C |
Be | 1100C |
GaTe | 540C |
N | 150W RF |
Al | Depends! Read recipe section! |