Photolithography - MJB4
Lithography - MJB4
Adopted from Process Flow for Tunnel Junctions and TLM Modified by NTS 01/23/13. Modified by NTS 140602, 150115, 150928.
Note: Sample must be cleaved and in sample holders to begin.
Prep
Setup hotplate and thermometers
Take photoresist out of the refrigerator because it needs to be at room temp
Crack PR fridge to let fumes get vacuumed up, then open.
AZ 5214E-IR
Check MJB4 lamp power
Check the power unit is working and on CH1/365
Insert empty mask holder, raise the stage, and expose for 10-20 seconds
Look at the output power listed on the power unit, it should be ~7.5 mW/cm^2
If it is not, you need to email Mertech to calibrate the lamp.
Power less than 6.5 can ruin your litho. 6.5-7.5 is not ideal but OK for process dummies.
Solvent Clean (ACE, IPA, DI: 2 min)
Create a clean workstation before beginning.
Spray bench with IPA if needed, then wipe clean
Do this by getting clean wipes from the dispenser and lay them down anywhere you plan to work.
For Regrowth, layer wipes from back to front, and right to left, and go up to/over lip of bench
Setup solvent beakers and dipper basket
Always fill each subsequent beaker higher than the previous
Note: The dipper baskets are labeled. Only use those that are appropriate for your immediate task. Otherwise, keep them in the bin and close the bin to prevent any contamination. Note: Gentle agitation during cleaning process is recommended.
Acetone: 2 min
2-propanol: 2 min
Go get UHV foil for boat
Scrap UHV foil is ideal, never trust other person’s boat
DI water cascade rinse: 2 min
Make foil boat
N2 blow dry
Put samples in boat
Dehydration bake
150C oven: 5 min
Let Samples cool in boat
Setup hotplate for prebake with thermometers
Set hotplate to prebake temperature
Thermometers are setup up in a particular way. Short silver is at NE position, tall silver is NW, tall black is SW, short black is SE. Thermometers should be on flat foil and positioned so that the edges of the hotplate are tangential to the thermometer. All thermometers should be rotated so that the dial is in normal, easy reading position.
Clean up acetone and IPA beakers
Spin PR
Spinner Prep
Make sure spinner chuck is clean
Our own 0.5” chuck should be clean
Spray wipe with acetone and wipe chuck off, swap gloves, repeat
Don’t get solvent on inside
Spray wipe with IPA and wipe chuck off
N2 dry chuck
If chuck is not at spinner you want it at, assume it’s dirty
Make sure you have pipettes for PR
Note: Set spinner to almost max time, always use your timer!
Check the spinner by placing a junk wafer piece or non-critical sample on it. Make sure it spins correctly and at the correct RPM (4k) before placing on a critical sample or PR.
Spin on resist (AZ 5214E-IR, 4k rpm, 30 sec)
Check Sample 1 and N2 blow
Put Sample 1 on chuck and center
Check centering and RPM
Note: When putting PR onto the wafer, try not to put on so much that it leaks off the sides because this can get into the vacuum and mess things up. Also, make sure that there is an adequate amount, as to completely coat the wafer with an even amount. If the integer number of drops you must apply is too little (too much) for the sample, wait for it to spread a little before spinning (spin immediately after application)
Put PR (AZ 5214E-IR) in pipette carefully and drop a few drops into the PR container (to eliminate air)
Drop on PR, no bubbles, and Spin
4k rpm
30 sec
Check wafer when done. No streaks.
Repeat as needed
When done, empty pipette into PR container and return to fridge
Clean chuck as above
Clean spinner bench
Pre-Bake
(evaporate solvents before lithography)
100C hotplate: 1 min
Let samples cool in boat while you prep aligner
Check that mask is clean
Set hotplate to IR bake temperature
Exposures
First exposure (Soft contact, 11 sec)
Choose stage for your sample
Prepare height of aligner stage to be close to (but lower than) what you will want with the sample.
Put Mask onto holder, ~1 mm from edge, turn on vacuum, slide into aligner, and tighten screws. Only touch the mask corners and edges.
Note: When placing the mask onto the holder, place it so that the chrome side of the mask faces away from the holder. (Brown side faces sample after placement.) Keep your hand under the mask when moving the loaded mask holder
Load sample onto stage and lower stage a few turns
While looking from the front of the aligner at the sample (between the mask and stage), rotate contact lever, and then raise the stage til it looks like the sample and its reflection on the mask are about to touch.
Align features, then raise stage til contact.
The best way to tell contact is by looking (through the microscope) for rings to appear at a corner of the sample or an edge
Can also look for a brightening of the mask metal, meaning there’s pressure on the mask
Whichever corner or edge displays contact first on your first sample will probably be the corner/edge to display contact first for the other samples.
Note: Will feel a tiny increase in resistance of the knob.
Parameters
Align + expose
Soft contact (we do this now to avoid sample movement)
11 sec on MJB4 contact aligner
Note: Do not face the aligner when the wafer is being exposed. Also, make sure to move your other samples away from the aligner, as not to ruin them from scattered UV light. Usually this means putting a toolbox or bin between them and the aligner.
Lower the stage using the front knob about four numbers (knob number increases)
Center the stage rotation knob then Rotate contact lever
Unload sample and Repeat as needed
IR bake
(Image Reversal Bake)
115 C hotplate: 115 sec (1 min 55 sec)
Note: Very important step, do not over/under bake
Remove mask from aligner
Second exposure
(Flood exposure = w/o mask)
60 sec flood exposure on MJB4 aligner
Do not use hard contact (can’t select for flood exposure)
Note: Only put one wafer on at a time so as not to ruin wafers with scattered UV light.
Develop and Inspect
30 sec in MF-26A followed immediately by DI cascade rinse: 2min
Note: Gentle agitation the whole time. Note: Does not stop developing after you take the wafer dipper out of the developer solution, make sure to immediately dunk into DI water and rinse.
N2 blow dry
Put stuff away and clean up bench
Inspect Pattern
The key parameter for this process is the IR bake. If the pattern looks like features are way too big (overdeveloped), then the sample was under baked. If the features look too small (underdeveloped), then it was over baked. Also, too long of an Exposure 1 can lead to underdev, and too short can lead to overdev.