GaP cell on Echo

GaP cell on Echo

  • GaP cell was re-installed Fall 2024

GaP set points:

 

GaP Base Temp

GaP Base Output

GaP Tip Temp

GaP Tip Output

 

GaP Base Temp

GaP Base Output

GaP Tip Temp

GaP Tip Output

Idle (manual)

~800

15.5

~658

0.0

Growth (manual)

~1060-1070

22.8

~866

16.8

Outgas (manual)

~1100

23.5

~900

19.3

  • GaP Base TC makes the sub TC go crazy, so I unplug the sub TC if I need to check the GaP Base Temp but for the most part I just assume that if the tip is stable, so is the base

  • Eurotherm PIDs have not been tuned for this cell yet.

 

Valve range: 0-200 mils in its current config

 

Notes on valve: Currently the valve faults pretty frequently when trying to open, my guess is that closing the shutter actually changes the temperature of the tip and results in it getting stuck, so during the wait time I tend to leave the shutter open, once its opening and closing for STOs it typically doesn’t get stuck but I am not sure how that’ll work during the junk growth where the close times are a little longer. I am going to tune the PIDs for the Tip and hopefully that should fix it, but I haven’t had a chance to. 

 

Using the GaP cell to grow: Heating up the GaP cell destroys the background of the BFG, so I wait to heat it up until after I do my IIIs and As BEPs. Once I am done growing, I idle the cell and rehome the valve.

 

Ramping up cell (ex. From idle to growth):

! Manually open valve to ~ 80 first

 

eval GaP_OP = 15.5

repeat 73

eval GaP_OP = GaP_OP + 0.1

setparameter Output GaPBase=GaP_OP

wait 00:00:30

er

 

eval GaP_OP = 0

repeat 168

eval GaP_OP = GaP_OP + 0.1

setparameter Output GaPTip=GaP_OP

wait 00:00:10

er

 

open GaP

! Manually open valve to ~ 80 first

 

eval GaP_OP = 15.5

repeat 73

eval GaP_OP = GaP_OP + 0.1

setparameter Output GaPBase=GaP_OP

wait 00:00:30

er

 

eval GaP_OP = 0

repeat 168

eval GaP_OP = GaP_OP + 0.1

setparameter Output GaPTip=GaP_OP

wait 00:00:10

er

 

open GaP

Ramping down cell (ex. From growth to idle):

! Valve opened to 90 and shutter already open

 

! set GaPTip setpt to 600C and ramped down in auto

! once at zero power put back in manual

 

! wait fot GaPTip to bottom out before running recipe

waituntil (815 < GaPTip < 820) 10

 

eval GaP_OP = 22.8

repeat 73

eval GaP_OP = GaP_OP - 0.1

setparameter Output GaPBase=GaP_OP

wait 00:00:30

er

 

waitop (once Tip and Base are stable rehome valve)

! Valve opened to 90 and shutter already open

 

! set GaPTip setpt to 600C and ramped down in auto

! once at zero power put back in manual

 

! wait fot GaPTip to bottom out before running recipe

waituntil (815 < GaPTip < 820) 10

 

eval GaP_OP = 22.8

repeat 73

eval GaP_OP = GaP_OP - 0.1

setparameter Output GaPBase=GaP_OP

wait 00:00:30

er

 

waitop (once Tip and Base are stable rehome valve)

Last time I outgassed the cell: 

  1. Basically just ran an STO a few times since everything else had already been outgassed and we are pretty sure the cell is leaking

eval t_GaPValve = 180

eval t_GaPValve_Step = 10

 

open GaP

! calculate

repeat 3

  repeat 2

    open Ga

    wait 20

    close Ga

    wait 25

    close GaP

    wait 2

    eval mBG=bf

    t GaPValve = t_GaPValve

    wait 2

    open GaP

    wait 40

    eval mP=bf

    eval BEP = mP - mBG

    writefile (GaPFlux; t_GaPValve, mBG, mP, BEP)

    t GaPValve 0

   

  er

  eval t_GaPValve = t_GaPValve + t_GaPValve_Step

er

eval t_GaPValve = 180

eval t_GaPValve_Step = 10

 

open GaP

! calculate

repeat 3

  repeat 2

    open Ga

    wait 20

    close Ga

    wait 25

    close GaP

    wait 2

    eval mBG=bf

    t GaPValve = t_GaPValve

    wait 2

    open GaP

    wait 40

    eval mP=bf

    eval BEP = mP - mBG

    writefile (GaPFlux; t_GaPValve, mBG, mP, BEP)

    t GaPValve 0

   

  er

  eval t_GaPValve = t_GaPValve + t_GaPValve_Step

er

  1. Outgassed the shutter by homing the valve and shuttering with long closed periods like we do in junk growth with Arsenic