Located in Share > library > MBE-manuals > Commercial > Veeco - Valved Cracker As_500

Equipment and Notes

Equipment

  • Wrench, Box End - 1/4” x 5/16”

  • Veeco Dial Torque Wrench and Socket (1/4”, 6 pt.)

  • Gasket – 6” Silver

  • Gasket plate - buy new from veeco

  • Moly bolts for sealing ring x 16 - buy new from veeco

Notes

  • Make sure valve is open! (Ensures crucible is vented and pumped down)

  • Can reload on the system

Open and Load

  • Purge the water lines with N2 over several hours &/or cap to prevent H2O presence.

  • Open needle valve if not already open.

  • Vent BOS

  • Remove the top of the sublimator and put aside (in clean foil)

  • Leave 6" gasket on to protect knife edge during crucible opening and loading

image-20240603-183627.png
As sublimator open as shipped with top removed

Note: New As cell's are shipped without a crucible gasket plate installed

  • Remove the sealing ring using a UHV clean torque wrench and place in clean UHV clean foil (save bolts in clean UHV foil boat)

image-20240603-183708.png
Removing bolts with torque wrench

Note: The bolts will not be re-used except if the cracker has just been sent from Veeco without a gasket plate installed (e.g. brand new or re-built)

  • Remove gasket plate (we do not reuse this) with UHV pliers on tabs (This can be thrown away)

image-20240603-183736.png
Sealing ring
image-20240603-183751.png
Empty crucible
  • P2 removes As bottle(s) from outer wrapping and cuts open last bag

  • P1 opens As bottle(s) and carefully pours contents into crucible

Do not get any As anywhere other than in the crucible! Especially in the sublimator but outside the crucible-- don't do that!

image-20240603-183842.png
Loading As chunks
  • P1 hands closed bottle back to P2 who seals it in a bag and puts aside

  • Repeat for second bottle until unable to continue filling crucible (~ 650g loaded on Echo 12/11/20)

image-20240603-183904.png
1 kg, 2 bottles loaded

Close

  • P1 carefully wipes inner crucible knife edge (and puts on new clean glove(s), like after any knife edge wiping)

  • P2 cuts open gasket plate bag with clean scissors

  • P1 uses clean gloves to remove gasket plate and insert into crucible with tabs aligned

image-20240603-183956.png
Gasket installed
  • P2 unwraps sealing ring for P1 to grab

  • P1 replaces sealing ring while P2 makes strong UHV foil boat for new moly bolts

Note: Unless we re-use the bolts, which is only if the cracker has just been sent from Veeco without a gasket plate installed (e.g. brand new or re-built)

  • P2 opens new moly bolts bag and pours them into strong UHV foil boat. P2 will hold this boat over the open UHV tool container in case it rips.

  • P1 inserts all bolts, finger tight

Sealing ring installed and secured by top & bottom bolts
Installing and tightening bolts by hand
  • P1 uses clean socket and veeco dial torque torque wrench to tighten bolts, sequentially 30 in/lbs, 60 in/lbs, 90 in/lbs 3x each

Torquing bolts to spec
  • P1 uses clean gasket pullers to remove 6" gasket from the top of the sublimator and wipes knife edges

  • P2 cuts new 6" silver gasket, P1 puts gasket on evaporator knife edge

  • P2 unwraps top of sublimator, P1 replaces top of evaporator, P3 closes N2 line

  • P1 replaces and tightens bolts

  • Prior to pump down, continuity check the cell