Ion Pump Feedthrough Change
Changing from old-style feedthroughs with ceramic protective coating to Gamma Vacuum feedthrough, using feedthrough conversion kit HVFT, SC (SHV-10) MINI CF, Part # 310002
Less fragile, susceptible to leaking
Currently installed on Auger ion pump and Auger buffer ion pump
Original ceramic feedthrough on ion pump
Procedure
Unblank ion pump
Remove original feedthrough
Unpackage and assemble new feedthrough
Insert new feedthrough into feedthrough port on ion pump, so that end of feedthrough intersects with copper connector piece.
Use clean gloves or ultraclean tweezers when touching inside of ion pump
Using ultraclean wrench & clean gloves or ultraclean tweezers, tighten nut holding feedthrough in place
Connect to gamma vacuum ion pump controller using safeconn extension cable.