MBE Echo System
System Echo
System Echo: Sources and Rocket Ports
Source PID Values
Reference: Adjusting Source Cell PID Values
'Low Temperature / Default PID values !'
Source Cell | Pb | ti | td | Target Temperature (C) | Ramp Rate |
|---|---|---|---|---|---|
InTip Default | 300 | 50 | 1 | 350 | 10 |
InBase Default | 250 | 75 | 3 | 200 | 10 |
GaTip Default | 200 | 75 | 5 | 350 | 10 |
GaBase Default | 200 | 190 | 5 | 200 | 10 |
BiTip | 57 | 37 | 2 | 600 | 5 |
BiBase | 47.8 | 55 | 9 | 550 | 5 |
La | 150 | 75 | 5 | 1000 | 5 |
Lu | 150 | 75 | 5 | 700 | 5 |
Si SUSI | 75 | 96 | 10 | 800 | 50 |
Be | 620 | 80 | 2 | 350 | 100 |
AsSub | 125 | 150 | 35 | 345 | 100 |
AsCracker | 250 | 70 | 5 | 850 | 100 |
Substrate | 22.3 | 68 | 11 | 200 | 60 |
BufferHTS | 128 | 158 | 26 | 200 | 30 |
Note: Only the first (low temperature) set of PID values can be read or written through AMBER. If gain scheduling is enabled, the second (high temperature) set of PID values must be read and written directly from the Eurotherm.
'High Temperature PID values (Gain Scheduling must be enabled) !'
Source Cell | Pb2 | ti2 | td2 | Target Temperature (C) | G.SP (C) |
|---|---|---|---|---|---|
InTip | 41.8 | 9 | 2 | 900 | 700 |
InBase | 45 | 17 | 5 | 750 | 550 |
GaTip | 41.8 | 9 | 2 | 1000 | 700 |
GaBase | 45 | 17 | 5 | 850 | 550 |
Si SUSI | N/A | N/A | N/A | N/A | 1400 |
Be | 470 | 12 | 2 | ~1000 | 700 |
Substrate | 100 | 50 | 2 | 700 | 350 |
BufferHTS | 110 | 100 | 5 | 700 | 30 |
Cell | P | I | D | Ramp |
|---|---|---|---|---|
LC Lamp | 106 | 38 | 6 | 60 |
Al | 300 | 150 | 5 | 5 |