System Echo Idle

System Echo Idle

Source IDLE Conditions!

First pass by SRB, updated EMK 5/4/2011
As Sublimator detail added by SDS 5/20/2015

Temperatures:

  • Al: Before melting 1.7% manual power, After melting 800C (with backup UPS in-line)

  • AsCrk: 850C

  • AsSub: 320C

  • InTip: 350C

  • InBase: currently OFF

  • GaTip: 350C

  • GaBase: currently OFF

  • La: 1000 C

  • Gd: 600 C

  • Lu: 700 C

  • BiTip: 310 C

  • BiBase: 300 C

  • Si: 350C

  • Be: 350C

  • Substrate: 200C or 4.5%

  • HTS: 10C or 0%

  • LCLamp: 10C or 0%

  • Heated Viewport 400C

CAR:

  • Beam Flux gauge on

  • In Transfer Position

As Sublimator:

  • When not growing with As, set the As Eurotherm in Auto mode at 357C. This is approximately 29.5% power. This setting may change depending on the needs of the various growers.

  • During growth, put the As Sub Eurotherm in Manual mode.

    • Standard Manual mode setting is 29.5%. This may change depending on requirements of the various growers.

PID values for Eurotherms (Pt, Ti, Td):

  • LC Lamp: 106, 38, 6