Pulling Shutter Wheel
This page describes how to remove the shutter wheel for both Echo and Bravo
Last updated: 06/10/2015 - SDM
Preparation
Cool down sources or bring to idle?
need older student to finish
Check that the shutters are in manual operation and closed
Remove Pyrometer
Remove set screw
Pull entire unit off machine
Disconnect the pyrometer's power cord and set it somewhere safe to avoid damage to its lens
Remove shutter rods (assuming shutters are taped)
Disconnect pneumatic power cords
x2 bottom right
x1 on top
Loosen set screws for connecting shutter rod to chamber (1/8" Allen)
1 set screw for each side and bottom. 2 set screws for top
Pulling the wheel
Remove shutter wheel
2 people pull
1 person to support and make sure nothing gets caught
Turn off the 3 heated view ports (i.e. slowly ramp to zero)
The 2 large blue/white side view port controllers are ramped manually at a rate of 50 degC/15 min. Keep (slowly) ramping these for the rest of this procedure
The 1 small control box for the backside viewport ramps at an automatic rate -- go ahead and ramp it all the way to zero
Disconnect source material water supplies and returns
As (Arsenic)
HT Bi (high-temperature bismuth)
DD
Groups IIIs