Piranha Clean
wearing order
- double general gloves+1 transparent plastic gloves
- cloth
- facial mask
- blue gloves
- make sure there is no holes and everything is good
Process
- test acid solution aspirator
- triple risen quarts container
- 1) 1 container with hydrogen peroxide(30%) 2) 1 container with sulfuric acid (96%) ratio 1:2=peroxide:Acid
- Combine together(peroxide pour into acid) -wait 5mins
- Put wafers in soluiton for 8mins
- QRD 5times→ Triple risen
- SRD
One wafer -200ml Hydronperoxide+400ml sulfuric Acid if using holder in hood
Take off everthing when leave acid hood!!