CHA 1

CHA 1

CHA#1 Procedure

Written by NTS, Jan 2013 from notes in my notebook. Last modified by AFB 20160930

 Standard Operating Procedures (MERTech) for CHA 1

Loading

  • Have clean samples or clean while venting

    • Note: If someone is before you, ask them to leave it vented for your load. Be ready to load when they are done.

  • Check Xtal life before loading sources and samples

  • Check Power is off

    • IG’s off

    • Power sources off

  • Vent, takes about 12 minutes

  • Go to acid hood, prepare and use surface cleaning solution

    • HCL:DI (1:10, 40:120 mL)

    • Note: Never approach acid bench w/out gear on. Can drop in wipes.

    • Suit up with inspected apron, mask, and proper sized gloves

    • Setup wipes now and put sample in basket

    • Pour DI into rinse beaker (400 ml)

    • Pour 120 ml DI into 400 ml etch beaker (want a ~1:10 ratio)

    • Pour 40 ml HCl into 100ml beaker (HCl is already 1:3 in bottles)

    • Pour acid into DI and quadruple rinse 100 ml beaker

    • Dip samples for 15 sec in solution (precision is not needed here)

    • Cascade rinse for 30 sec in DI

    • Blow dry with N2

    • Clean beakers and workspace

      • Our acid solution is dilute so you don’t need aspirator

    • This part takes about fifteen minutes, so start it once you vent CHA #1

    • Note: This takes 20-45 minutes depending on number of samples

    • Carefully take off acid gloves, mask, and apron

    • Take samples immediately to CHA #1

  • Mount clean samples on holders. There is a clip set for the Bank group.

  • Prep chamber

    • Inspect bottom of shutter and hearth

      • Look for scraps on hearth or flaking on shutter bottom

    • To clean shutter, double glove, wipe with hand, and vacuum hearth; reglove

    • Clean (vacuum) chamber and rim where bell jar seals if needed

  • Load samples

  • Open shutter and load sources

    • Make sure you can see source both ways

      • Using all Si wafer mirrors

      • Using just back mirror

    • Close shutter when done

  • Check rotation with samples loaded before you pump down

  • Pump down, takes about 30 minutes

    • Don’t leave machine til pump crossover

    • Turn on IG (IG2) after crossover, when P < 1e-3/4 T

  • Check LN2 level using straw

    • Add LN2 if needed

    • Acquired in back chase

Evaporation Prep

  • Dep controller

    •  

      • Should be on the main screen

    • Press menu, go to general parameter, press menu again

    • Set process number by material (chart on breaker box)

    • Move cursor, press menu, go to operate, and press menu

    • Check that process loaded

    • Zero the thickness

    • Note: If the controller seems stuck, or can’t get to process number slot, press reset

  • Wait until chamber gets to < 5e-6 or < 6e-6

  • Choose first crucible

  • Make sure you have enough time on LabAccess before activating power for large stacks

    • If your stack needs more than 2 or 3 hours, consider LabAccess time before each run.

  • Turn on main breaker

  • Turn on Gun control

  • Turn on HV control

    • Hold up until breaker clicks twice, about a full second

  • Turn on rotation

    • If it makes a clicking sound in the chamber, it is not working right so turn it off

Evaporation

  • Press start. Tool auto ramps to 25%, Soak 1.

  • At Soak 1, you can take it to manual power. Press manual power (5)

    • Note: When you can see the ebeam, make sure it is not hitting the hearth or crucible, ever!

  • Ramp up to desired power at < 0.1% per sec

    • Note: Controller is finicky

  • Desired power level is determined by you

    • From other users, papers, past runs, etc.

  • When within a few percent open the shutter so you can guide the ebeam

    • Use beam sweep control 1

      • Ampl knobs: CW increases

      • Freq knobs: CCW increases

    • Metals other than gold need some motion

  • It’ll start depositing, so control dep rate.

    • At first, keep the dep rate under 0.5 A/s

    • Don’t ever exceed 1 A/s, metals can sputter, especially Ge

  • When done, or nearly so, always lower dep rate to 0.5 A/s or lower, I suggest ~0.1 A/s

    • 0.5 A/s can still cause as 4x spike in dep rate for a second

    • Ricardo suggests ramping down power/dep rate rate towards the end of your run, to avoid splashing material on shutter (and to get nice, slow layer)

  • Close shutter when done

  • We must now use an automated ramp down Idle Ramp that takes 5 minutes except for Al and Cr

  • Press Manual Power to start Idle Ramp down, will go down to 0%

    • Old way: Ramp down at < 0.1% per second. When you get to < 20%, can go much faster

  • When done, displays Idle and End of Process on controller at 0%

    • This will be checked, it can’t be left in manual

  • (Push in stop when power is at 0.0%. Not ever otherwise!)?

  • Controller must be left in Idle and End of Process or Ready state for next user, not manual

  • Turn off GC and HV

Switch Source

  • Switch pocket (source)

  • Chamber ready when interlock at hearth control goes on

  • Press reset on dep controller

  • See “Evaporation Prep” and “Evaporation”

Finishing

  • When done with GC and HV off

  • Press reset and zero on the dep controller

  • Turn off rotation

  • Turn off power

    • Turn off circuit breaker

    • Turn off IGs

  • Press Standby, wait for gate valve, press Vent, takes about 12 minutes

  • Remove samples and source

  • Can log off labaccess for pump down if no one is after you

  • If there is someone after you, you may not have to pump down chamber after use.

  • To pump down after use

    • Push switch to lower and wait for bell to lower and chamber to pump to mT range.

    • The pump might start to cross over. That’s when you set the system to standby.

    • Change switch to standby

  • Fill out log

Beam Sweep Notes

  • Ampl knobs: CW increases

  • Freq knobs: CCW increases

  • Beam sweeping skill takes some practice.

  • The beam when it is centered will be a little circle. If it is too high it will be ovalish. If it is too low it will look like a v thing (upside or sideways maybe).

  • It is good to raster a little bit of gold in center of source, but not too small a spot.

  • Other metals need rastering over most of surface

Stacks

  • Laser Top: Ti/Pt/Au : 100/100-300/2000 || Ti/Au : 100/1000

  • Laser Bottom: Au/Sn/Ni/Au : 100/100/100/1000