FEI Quanta 650 ESEM
FEI Quanta 650 ESEM Procedure
NTS, 06/06/2014, updated NTS 2014MMDD
[Note: This tool is located at TMI in FNT 3.110 on Main Campus.]
[Note: This tool has a lot of free time. The reservation horizon for this tool is ~one week.]
Prep
Fill in Log when you enter.
Right computer & screen is SEM –do not touch tower.
Left computer & screen saves images and does EDS. Can use USB and internet.
If mouse gets janky, press button on medium whitish box that says 1 or 2 to switch mouse control
Start at Vacuum and Beam control tab on right of screen, Sun icon, left tab (status tab)
Check Pressures and current
Mouse over green vase icon at bottom
IG1: gun chamber?
IGP: ion pump
Emission current
Load/Unload Sample
Press Vent on Vacuum frame of tab
Vase turns grey when done (2-3 min)
Mount sample/s on stub/s.
No sample size restriction as long as it can be mounted on stub (and stage)
They provide DS carbon tape and gloves
They provide stubs, maybe
When vented, gently pull out door using aluminum bar
Put stubs with samples into stage
Insert stub/s fully into stage
Stage stub holes have numbers, remember where you mount samples
Close door fully
Click Pump button and press door closed for a few seconds
Vase is orange, wait til it’s green, P < 2e-4 T (2-3 min)
On Vacuum frame of status tab, we will want high vacuum
Low vacuum is for non-conductive samples and uses water vapor to see. E.g. samples that can’t be coated because of EDX
E-SEM is environmental SEM for wet samples
When on high vac ignore chamber pressure in vacuum frame near top. Look at chamber pressure at bottom.
Software Prep
Adjust Spot size and High voltage (in Column frame)
Spot size depends on sample size, ranges from 1 to 7.
Smaller spot size is less current and v.v.
Usually 1 is not good enough, 7 is too much.
3 or 4 is recommended
High voltage is 1-30 kV
Depends on sample resistance to damage, sensitivity
These things can be changed after turning beam on.
10. There are multiple detectors. Default is to have Imaging detector and Chamber camera on
Move sample into position
Vertical working distance is 10 mm, is marked
Stage tab at top right, Stage/gear icon, second tab
Go to Coordinates tab
Double click Position 1 for stage to go to stub spot 1 (or Position X for spot X)
Now manually raise stage
Click Chamber detector screen
Click and hold mouse wheel, a yellow bar appears
Drag up and stage will move up and v.v.
Get sample into position.
Turn on Beam
First tab, click Beam On
Click first detector window and unpause
Pause/unpause button at top middle
Joystick has X, Y, rotation control
Can also use mouse by pressing and holding mouse wheel again
Make sure you have correct detector selected!
Can also double left click to move-to
Hard panel control of Mag and Focus, Brightness and Contrast
Autobrightness and contrast are OK. Don’t use autofocus or autostigmation.
Note: If no image when Beam is turned on, adjust brightness and contrast first. Next, check detector selection is correct. Go to Detectors dropdown at top left. We want ETD (SE) selected.
Press F5 to go from small screen to full screen (and back)
Adjust scanning dwell time
Ranges from 300 ns (default) to 1 ms
300 ns is easiest for focus and alignment.
Tortoise and hare buttons: slower and faster
Focus and alignment (minor)
Shouliang calibrates it weekly, we just have to do focus and stigmation adjustment on each sample like normal
Zoom in, focus decently, adjust stigmation, repeat as needed.
Streaking means stigmation is off.
Try to focus to a non-streaking position before adjusting stigmation
Can use Reduced area to do focus and get faster response
Helpful at higher mag too
Click Reduced area button at top left. Click again to go back.
Can reduce/adjust scan speed for higher quality pictures.
Use
Camera button at top center pauses image, save as pops up
Create own folder your first time
Files are saved on left computer
Unpause to continue
Ruler button at top right to measure features
Press delete to delete (FILO deletion order)
Different filter modes: Live, Average, Integrate
Live is normal
A filter mode from previous user could confuse you
Finish Use
Pause image for imaging detector.
Make sure Chamber detector is unpaused, to watch stage move
Click Vent, auto-lowers stage for you
Unload sample
Close door
Pump down
Finish log info
EDX
Open Esprit software on left computer and login
Runs very similar to 2nd floor one (Steps 38, 39, 41-44)
Get sample into position
Use spot size of 4 maybe 5 for EDX
27b2. Image quality will go down but EDX results might be better
Want spectrometer input counts to be > 1 kcps
Gun kV needs to be large enough, might need to change/increase
Go to Spectra mode and acquire spectrum
Select elements as usual, or can click auto button and deselect wrong elements.
Higher energy peaks have lower detection efficiency.
Save spectrum as normal if desired, spreadsheet or graphic
Click Objects to do mapping > Mapping Mode
Software chooses which peaks to use
For mapping, you can adjust this by adjusting peaks on elements tab
Use as normal
Point scans: scans/acquires spectrum at crosshair (no dimensions just counts vs. energy)
Line scan: scans/acquires along a line you draw and can adjust (one dimension, is composite of elements, looks like a Dektak overlap)
Mapping: acquires a false color map in 2D
Hypermap: makes a spectrum at all points (very slow, select small area)
Turn off/close when done!