AFM
Atomic Force Microscopy (AFM)
Introduction
Background Information:
Brief AFM Principles of Operation
Adorable AFM Flash Animation
Training
Basic training with another LASE member first
Certification with Marylene
Additional certification with Marylene for tip change
AFM 1 Measurement Procedure
Log into AFM1
Turn on Camera
Turn on old screen on the left side of the desk
Turn on illumination source by twisting knob on desktop tower
Put sample on the stage and turn on vacuum
Note: the stage only moves halfway under, so samples might need to be placed farther forward. Also note that the stage can be rotated
Open the software on the desktop called V614r1
Turn on the laser
Click on the real-time button that looks like a gold laser at the top left side of the window
Open the following windows by double clicking on their icons (left side of screen):
"Meter"
"Navigate"
"Scan single"
Align the system
Click "Locate Tip"
This focuses on the tip
Adjust z and the focus/zoom if needed
Press "OK" if the tip is sufficiently focused
Move sample under the laser using the XY stage control buttons
Decide what you want to focus on when bringing the tip near the sample. Pick between "Tip reflection" or "Surface". For samples that have a reflective surface, you want to focus on "Tip reflection"
Adjust the z height
Start with a quick speed, like 100, when lowering until some of the features come into view
Adjust speed to a lower value, like 50, and lower until the tip or the surface come into better focus
Lower the speed again, like 25, until closer
Lower speed, like 10, for a fine adjustment
Align laser using the laser alignment knobs until it is centered on the tip (might want to dim the sight source box power to make it easier to see)
Align the photodetector using the the photodetector knob until the vertical deflection is below 0.1 V
Tune the system
Click the tuning fork button
Change the peak offset to 5%
Click 'Auto tune'
Look for a Gaussian amplitude graph and sigmonic (i.e. backward S) for the phase. If it looks bad, the tip is probably bad
Perform the scan
Click on the "Real Time 1 - single scan" window
Click the "Engage" button
Set up the scan
Change the scan size to the size you care about, e.g. 30 um
Make sure the scan is in 'tapping' mode
The scan is working when the 'trace' and 'retrace' graphs are aligned
Adjust the amplitude setpoint to align the curves.
Note: the force of the cantilever on the samples is inversely proportional to the amplitude setpoint. This means the curves will be better aligned for a lower amplitude (i.e. higher force) because the surface is being traced better
Capturing images
Under the "Real Time tab", choose "capture filename"
Choose saved location in your directory, give the file a useful name, and hit "ok"
In Real Time 1 single scan window 'Capture' will show and there will be a 'capture: next' in the bottom left side of the software window
Image will start capturing once arrow reaches top or bottom of frame. To speed this up, hit "frame up" or "frame down" button
Bottom right of screen reads "Capture Next" when waiting for arrow to reach end of frame, "Capture on" during capture, and "Capture done" once complete
Do not adjust the amplitude error or other scan parameters during capture! This will end capturing of your frame. Adjusting the scale bar in any scan window is ok.
Ending scan
Change scan size to zero
When stabilized (trace and retrace are flat and constant), hit "withdraw" or "disengage" button. I like to hit this twice to be safe.
Raise AFM tip with z motor control up and down arrows
Make sure the speed is set to slow at first, then you can change to fast mode once you are confident you are well clear of the surface
Make sure you press the up arrow!!!
Once tip is raised, move stage out with xy.
Do not hit the hard stop at the edge of the marble
Replace sample if necessary.
Image processing
Flatten (steamroller button): smooths out image (gets rid of contrast artifacts from tip crossing large features). Usually 1st order flatten is appropriate.
Roughness (ruler button): measures surface roughness of image or of area. Rq is the typically reported value.
Step height (step button): measure step height of feature
Copy images from "capture" folder to z drive- can be accessed from computer outside of clean room with a thumb drive.
When finished at the AFM
Close all software windows (automatically happens if you close the main window)
Turn off the light control box
Turn off the tip computer monitor
Sign out of the lab access system
AFM 2 Measurement Procedure
Setup
Open software
Open workspace (if you don't have an existing workspace, copy Emily's "Tapping_2015.wks" file to your own directory). When copying someone's workspace, always make sure it's set up for tapping mode.
Click on navigate to activate workspace
Center your sample on the stage, turn on vacuum
Use xy arrows in software to move your sample under the laser beam, roughly in the area of interest
Focusing on Tip
Click "locate tip"
Use front knobs on AFM to move crosshair if not centered (see picture)
Use up and down arrows to focus on tip (use slowest speed). Confirm focus.
In the "Focus on:" drop down menu choose "tip reflection". This avoids crashing the tip into the surface when you bring the tip down. If your surface is not adequately reflective, you may have to choose to focus on the surface, but this should be avoided.
Use the z meter to lower the tip until the tip and the tip reflection overlap. While bringing the tip down, you will pass through a position where the surface is roughly in focus. This is ok. When the tip and tip reflection shadows start to approach each other, switch to the slowest speed. Keep your eyes on both the image and the physical tip in order to prevent crashing.
Open the "meter" window in your workspace
Decrease the illumination in your image and use the top knobs on the AFM to align the laser. The laser should be centered on the crosshair.
Use the side knobs on the AFM to center the detector. The dot on the screen should be red and near the center of the screen. The magnitude of the vertical deflection should be < 0.1. The intensity of the signal depends on the tip, but should be optimized. If the dot is grey and moving randomly, the laser is too far out of alignment, and cannot be detected by the detector.
Preparing for scan
Under "Focus on:" Choose surface
Move in xy to find features, if applicable. Use slowest speed.
Click on Tune (tuning fork icon)
Autotune- top graph should be symmetric bell curve, bottom graph should be backwards S shape (tuning graphs for good quality tip shown below). If tip needs to be changed, contact Emily, Hema, Stefano or Li Tao.
Close hood (note- always do this before engaging tip with surface)
Check scan parameters under "scan control"
Samples/line: 128 lowest, 512 for good quality. More samples = longer scan
Microscope mode: tapping
XY closed loop: off unless features > 200 nm expected
Scan size: 0
Aspect ratio: 1
Integral gain: 0.6
Proportional gain: 1.2
Choose scan windows- I typically choose height, amplitude error, and phase
Scan
Click on scan - height window
Click engage button. Software will check distance from surface and bring tip into contact with surface
Once red and blue (trace and retrace) lines are consistent, increase scan size to desired range for your scan. 5-10 um gives you a good idea of surface roughness for thin films. It's a good idea to scan several areas of the same sample.
Adjust amplitude as necessary under scan control to improve match between trace and retrace lines
Check "real times status" green status bar to make sure tip is not over or under extended. Should be < 30 in magnitude.
Capturing images
Choose "capture filename"
Choose saved location in your directory
Image will start capturing once arrow reaches top or bottom of frame. To speed this up, hit "frame up" or "frame down" button
Bottom right of screen reads "Capture Next" when waiting for arrow to reach end of frame, "Capture on" during capture, and "Capture done" once complete
Do not adjust the amplitude error or other scan parameters during capture! This will end capturing of your frame. Adjusting the scale bar in any scan window is ok.
Ending scan
Change scan size to zero
When stabilized, hit "disengage" button. I like to hit this twice to be safe.
Raise AFM hood so you can keep an eye on the tip while raising it
Raise AFM tip with up and down arrow. Make sure you press the up arrow.
Once tip is raised, move stage out with xy. Replace sample if necessary.
Image processing
Flatten (steamroller button): smooths out image (gets rid of contrast artifacts from tip crossing large features). Usually 1st order flatten is appropriate.
Roughness (ruler button): measures surface roughness of image or of area. Rq is the typically reported value.
Step height (step button): measure step height of feature
Copy images from "capture" folder to z drive- can be accessed from computer outside of clean room with a thumb drive.
Operating Procedure from Tutuc's group: File:Digital Instruments AFM.pdf
Tip Change Procedure
Use z positioner to move AFM tip a few cm above sample
Unscrew AFM head using side screw. Screw turns opposite way from anticipated.
Remove tip holder from AFM head, place on holding stage
Unclip tip and replace with new tip. Make sure new tip is not touching sides of metal channel
Replace tip holder on AFM head. Make sure dots line up- different spacings on each side
Holding AFM head, move laser so that it's partially covered by tip. Strategy: turn horizontal displacement knob the full way clockwise, then one turn back- then turn vertical displacement knob until you can see the shadow of the tip covering the laser dot. If the dot totally disappears, it is probably hitting part of the tip holder, not the actual tip.
Replace AFM head, tighten screw. Make sure it is firmly seated, so that it doesn't drop into place unexpectedly later
Locate and lower tip
Fine alignment of laser and detector
Make sure to autotune the tip
Should get vertical deflection below 0.1 and the RMS amplitude between 400 and 500 mV
Record tip change in log book.
Trouble shooting
Issue: curve shapes don't look right during autotune
Possible causes: Tip probably needs to be changed. To check quality of tip, choose "Compute Q". Q value should match range specified on tip box.
In order to change the AFM tip, you need an additional certification from Marylene. The extra tips are in Emily's green toolbox in the PL lab. Users certified to change the tip: Emily, Hema, Tanuj, Stefano
Issue: triangular artifacts during scan
Possible causes: bad tip, sample is wobbling on stage (can happen with larger samples), tip was not initially focused well
Disengage from surface
Check autotune
If autotune is ok, check sample position. Do not move the sample when the tip is lowered, even if not engaged with surface
If ok, try re-aligning laser and detector and re-engaging with surface
Issue: Error message during startup - software can't connect to stage or z positioner
Try restarting software
Try a different workspace
Tools --> Select Microscope --> Should be only one option, select that one.
Issue: Poor resolution
Change the tip! Order sharper tip if necessary.